Labmate NXT Clinical Chemistry Analyzer

Brand: Trivitron

FEATURES

  • Accurate & precise results with maintenance free photometer
  • Direct report printing on letter head with Labmate NXT through external printer
  • Ideal for laboratories with small and medium workload
  • Lamp saver mode to increase lamp life
  • Unlimited patient result storage
  • 4 level of control run
Compare

Description

OVERVIEW

Labmate NXT instrument is a high performance semi-automatic biochemistry analyser for routine chemistries, electrolytes and drug tests.

It is equipped with user friendly software with 8 modes of operation, ease of programming, performing the testing and reporting the patient results. The analyzer has 10,00,000 patient results storage, real time curve monitoring, 4 levels of quality control with unlimited storage and LJ chart for daily and monthly QC results, calibration data and graph storage.

The analyser comes with 7 inch color display, capacitive touch screen which allows user to migrate from one window to another without going through complex sequential operations. A Patient ID wise, Date wise and or Test wise collated report can be generated upon request.

This analyzer is intended to be used for In-vitro quantitative and qualitative determination of wide range of analytes in body fluids by the trained person, in controlled environment and continuously monitored operation.

The State-of-Art features, advanced optics, versatile analytical capabilities and 32 micro-litre flowcell making the analyzer a smart choice for all laboratories.

TECHNICAL SPECIFICATIONS

Optical Unit

  • 12V/ 20W Quartz Halogen Lamp.
  • Monochromatic and Bi-chromatic.
  • Range of absorbance is 0 – 3.5 Abs.
  • 6 filters + 2 Optional: 340nm, 405nm, 505nm, 546nm, 578nm, 630nm (Optional: 450nm & 700nm).
  • Filtration mechanism- Rotating (Filter wheel), using stepper motor.

Measuring Unit

  • 32 µl Quartz Flow cell
  • Peltier Temperature Control: 20°C-40°C in steps of 1°C Digitally controlled using Peltier module and sensor.
  • Patient result Collated by ID, date and Test.
  • Aspiration volume- 200-1000 µl
  • 7 inch Color Display with Capacitive Touch screen
  • In-built Incubator: 8 Positions for Standard Round Test Tubes
  • Fixed at 37°C ±0.5°C, digitally controlled using Silicone heater pad and sensor

Analysis Methods

  • End Point.
  • Fixed Time
  • Fixed Time with multi standards
  • Kinetic
  • Kinetic with Standard
  • Multi Standard
  • Absorbance

Additional information

Brand

Trivitron